题名螺钉F角对腰椎后路单节段固定稳定性的影响
其他题名Biomechanical analysis of posterior monosegmental fixation: effect of the angle of pedicle screw implanted
作者
发表日期2009-08-15
发表期刊医用生物力学   影响因子和分区
语种中文
原始文献类型学术期刊
关键词椎弓根 单节段 内固定 生物力学 循环载荷
摘要目的比较4组不同椎弓根螺钉置入方式对后路单节段固定生物力学稳定性的差异。方法将24具新鲜小牛胸腰椎标本(T11~L3)分为4组,用椎体楔行切除法在L1椎体上制作严重压缩性骨折模型,用4组不同进钉角度行后路单节段椎弓根钉内固定,对固定后的标本施加频率为1.0Hz的前屈/后伸、左/右侧屈和左/右旋转疲劳载荷各3000次,经脊柱三维运动测量系统测量正常、损伤、固定和疲劳后4种状态下固定节段(T13~L1)前屈/后伸、左/右侧屈和左/右旋转运动范围,将其标准化为稳定指数后比较四组内固定方式在4种状态下6个载荷方向上稳定性的差异。结果4组固定均能显著提高骨折模型在6个载荷方向上的稳定性(P<0.01),而且显著强于正常标本组(P<0.01),但四组间无显著性差异(P>0.05);疲劳试验后4组标本在6个载荷方向上的稳定性均小于各自疲劳试验前,但两者间无显著性差异(P>0.05),且4组间无显著性差异(P>0.05)。结论只要保证螺钉在椎弓根及椎体内且伤椎螺钉避开骨折区域,螺钉的F角大小不会影响单节段固定的即刻稳定性及疲劳后稳定性。
其他摘要Objective Compare the biomechanical properties of four groups with different posterior monosegmental fixation and make the evaluation.Method Twenty four fresh frozen of one-week-old calf spines(T11-L3)were divided into four groups following the introduction of incomplete compression fracture at the vertebral body of L1 by the method of excision.Four types of monosegmental fixation were applied to the four groups respectively to restore spinal stability.A cyclic loading were applied to the specimens at rate of 1.0 Hz up to 3000 cycles in flexion/extension,left/right lateral bending and left/right rotation respectively.Segmental stability(T13—L1)tests were then performed on the specimens in the condition of intact,injured,fixation respectively.Range of motion(ROM)in flexion/extension,left/right axial rotation,left/right lateral bending were determined by three dimensional laser scanner.Then the ROM was standardized into stability potential index(SPI)to compare the stability in six loading directions of four types of fixation in four conditions.Result All of the four groups with monosegmental fixation could significantly increase the stability in six motion directions when compared to their injury condition(P<0.01),and significant superior than that of intact condition(P<0.01),but there was no significant difference among the four groups(P>0.05).The stability in six motion directions after cyclic loading was inferior when compared to that of before cyclic loading respectively,but there was no significant difference in between(P>0.05),or among the four groups(P>0.05).Conclusions As long as it could ensure to implant the pedicle screw into pedicle and vertabra,and avoid the fractured areas in the affected segment,the angle of pedicle screw directed to superior or inferior endplate will not affect both the instant and post-fatigue stability.
资助项目浙江省温州医学院重大科研专项资助项目(QTJ06012);广东省自然科学基金项目(5300714)
ISSN1004-7220
卷号24期号:04页码:285-289
页数5
收录类别CNKI ; 万方 ; CSCD ; PKU ; 北大核心 ; ISTIC
学科领域医药、卫生 ; 基础医学
URL查看原文
CSCD记录号CSCD:3665881
引用统计
文献类型期刊论文
条目标识符https://kms.wmu.edu.cn/handle/3ETUA0LF/77479
专题附属第一医院
第二临床医学院、附属第二医院、育英儿童医院
数字化医学与智能技术研究院
作者单位
1.温州医学院附属第一医院;
2.温州医学院附属第二医院;
3.温州医学院数字化医学研究所
第一作者单位附属第一医院
第一作者的第一单位附属第一医院
推荐引用方式
GB/T 7714
朱旻宇,池永龙,吴立军,等. 螺钉F角对腰椎后路单节段固定稳定性的影响[J]. 医用生物力学,2009,24(04):285-289.
APA 朱旻宇, 池永龙, 吴立军, 黄其杉, 徐华梓, & 毛方敏. (2009). 螺钉F角对腰椎后路单节段固定稳定性的影响. 医用生物力学, 24(04), 285-289.
MLA 朱旻宇,et al."螺钉F角对腰椎后路单节段固定稳定性的影响".医用生物力学 24.04(2009):285-289.

条目包含的文件

下载所有文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
YISX200904009.CAJ(539KB)期刊论文出版稿开放获取CC BY-NC-SA浏览 下载
个性服务
查看访问统计
谷歌学术
谷歌学术中相似的文章
[朱旻宇]的文章
[池永龙]的文章
[吴立军]的文章
百度学术
百度学术中相似的文章
[朱旻宇]的文章
[池永龙]的文章
[吴立军]的文章
必应学术
必应学术中相似的文章
[朱旻宇]的文章
[池永龙]的文章
[吴立军]的文章
相关权益政策
暂无数据
收藏/分享
文件名: YISX200904009.CAJ
格式: CAJ VIEWER
此文件暂不支持浏览
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。