题名 | Influence of different implant designs on replacement of four teeth of the posterior free-end edentulism: Three-dimensional finite element analysis and clinic case validation |
作者 | |
发表日期 | 2023-08 |
发表期刊 | Annals of anatomy = Anatomischer Anzeiger : official organ of the Anatomische Gesellschaft 影响因子和分区 |
语种 | 英语 |
原始文献类型 | Journal Article |
关键词 | Finite element analysis Implants Mandible Prosthetic dentistry Stress |
其他关键词 | FIXED DENTAL PROSTHESES ; STRESS-DISTRIBUTION ; BONE ; NUMBER ; DENTISTRY ; DIAMETER ; SURVIVAL |
摘要 | With periodontal disease having an increasing incidence, mandibular free-end edentulism caused by periodontitis is clinically more common. Finite element analysis and clinical case reports were used to evaluate the influence of different designs on the load distribution of implant prosthesis in mandibular posterior free-end edentulism., A finite element model of a mandible with posterior free-end edentulism was established. Considering the implant position and selection of single crown repair or splint repair, four designs were conducted including model A: 3435 × 37(four-unit fixed bridge supported by three implants, implant positions were 34, 35, 37); model B: 34,35 × 37, (34: a single implant crown) (35 ×37: three-unit fixed bridge supported by two implants, implant positions were 35, 37); model C: 34 × 3637(four-unit fixed bridge supported by three implants, implant positions were 34, 36, 37); and model D: 34 × 36, 37(37: a single implant crown)(34 ×36: three-unit fixed bridge supported by two implants, implant positions were 34, 36). Stress distribution and the Von Mises stress value of the implants, the crown and the bone around the implants were analyzed at vertical and 45° inclined load., Stress in the cortical bone was mainly concentrated around the implant neck, and maximum Von Mises stress (MVMS) of the four models was 11.6-16.1 MPa at vertical load and 61.74-96.49 MPa at 45° inclined load. Stress in the cancellous bone was concentrated around the implant base, and MVMS of four models was 3.075-3.899 MPa at vertical load and 5.021-6.165 MPa at 45° inclined load. Stress of the restoration crowns was mainly concentrated in the connector of the bridge, and MVMS of four models was 23.38-26.28 MPa at vertical load and 53.14-56.35 MPa at 45° inclined load. Stress of the implant interface was mainly concentrated on the surface of the smaller implants near the bridge, and MVMS of four models was 21.12-33.25 MPa at vertical load and 83.73-138.7 MPa at 45° inclined load., There was favorable stress distribution of the four models at vertical load and 45° inclined load. Design of a three-unit fixed bridge combined with a partial crown may be an available option for devising patient treatment plans with mandibular free-end edentulism. |
资助项目 | Wenzhou Municipal Science and Technology Bureau Projects, China [Y20220378, ZY2022022]; Wenzhou Major Scientific and Technological Innovation Project [81271663]; National Natural Science Foundation of China [31471146, QTJ06012]; Wenzhou Medical University Scientific Development Foundation of Zhejiang, China [2022R413A044]; Zhejiang Province Science and Technology Plan Research and Xinmiao Talent Program, China; [Y2020594] |
出版者 | ELSEVIER GMBH |
ISSN | 0940-9602 |
EISSN | 1618-0402 |
卷号 | 249 |
DOI | 10.1016/j.aanat.2023.152111 |
页数 | 8 |
WOS类目 | Anatomy & Morphology |
WOS研究方向 | Anatomy & Morphology |
WOS记录号 | WOS:001056662900001 |
收录类别 | PUBMED ; SCIE ; SCOPUS |
在线发表日期 | 2023-06 |
URL | 查看原文 |
PubMed ID | 37247761 |
SCOPUSEID | 2-s2.0-85160807366 |
通讯作者地址 | [Ding, Xi]Department of Stomatology,The First Affiliated Hospital of Wenzhou Medical University,Zhejiang,Wenzhou,325015,China |
Scopus学科分类 | Anatomy;Developmental Biology |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | https://kms.wmu.edu.cn/handle/3ETUA0LF/180602 |
专题 | 其他_附属平阳医院(平阳县人民医院) 附属第一医院 数字化医学与智能技术研究院 |
通讯作者 | Ding, Xi |
作者单位 | 1.Department of Stomatology,Pingyang Hospital Affiliated of Wenzhou Medical University(The People's Hospital of Pingyang),Zhejiang,Wenzhou,325400,China; 2.Institute of Digitized Medicine and Intelligent Technology,Wenzhou Medical University,Zhejiang,Wenzhou,325015,China; 3.Department of Stomatology,The First Affiliated Hospital of Wenzhou Medical University,Zhejiang,Wenzhou,325015,China |
第一作者单位 | 其他_附属平阳医院(平阳县人民医院); 数字化医学与智能技术研究院 |
通讯作者单位 | 附属第一医院 |
第一作者的第一单位 | 其他_附属平阳医院(平阳县人民医院) |
推荐引用方式 GB/T 7714 | Ye, Zhangyan,Ye, Hao,Jiang, Zhengting,et al. Influence of different implant designs on replacement of four teeth of the posterior free-end edentulism: Three-dimensional finite element analysis and clinic case validation[J]. Annals of anatomy = Anatomischer Anzeiger : official organ of the Anatomische Gesellschaft,2023,249. |
APA | Ye, Zhangyan., Ye, Hao., Jiang, Zhengting., Wu, Suwen., Feng, Zhongqiang., ... & Wu, Lijun. (2023). Influence of different implant designs on replacement of four teeth of the posterior free-end edentulism: Three-dimensional finite element analysis and clinic case validation. Annals of anatomy = Anatomischer Anzeiger : official organ of the Anatomische Gesellschaft, 249. |
MLA | Ye, Zhangyan,et al."Influence of different implant designs on replacement of four teeth of the posterior free-end edentulism: Three-dimensional finite element analysis and clinic case validation".Annals of anatomy = Anatomischer Anzeiger : official organ of the Anatomische Gesellschaft 249(2023). |
条目包含的文件 | 条目无相关文件。 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论