科研成果详情

题名Anatomical and Biomechanical Stability of Single/Double Screw-Cancellous Bone Fixations of Regan-Morry Type III Ulnar Coronoid Fractures in Adults: CT Measurement and Finite Element Analysis
作者
发表日期2023-04
发表期刊Orthopaedic surgery   影响因子和分区
语种英语
原始文献类型Journal Article
关键词Anatomical and Biomechanical Stability Digital Orthopaedics Surgery Finite Element Method Smart Screw Internal Fixation Ulnar Coronoid Fracture
其他关键词PULLOUT STRENGTH ; RADIAL HEAD ; ELBOW
摘要At present, it is still uncertain whether single screw has the same stability as double screws in the treatment of ulnar coronal process basal fracture (Regan-Morry type III). So, we aimed to compare the pull-out force and anti-rotation torque of anterior single/double screw-cancellous bone fixation (aSSBF, aDSBF) in this fracture, and further study the influencing factors on anatomical and biomechanical stability of smart screw internal fixations., A total of 63 adult volunteers with no history of elbow injury underwent elbow CT scanning with associated three-dimensional reconstruction that enabled the measurements of bone density and fixed length of the proximal ulna and coronoid. The models of coronal process basal fracture, aSSBF and aDSBF, were developed and validated. Using the finite element model test, the sensitivity analysis of pull-out force and rotational torque was carried out., The pull-out force of aSSBF model was positively correlated with the density of the cancellous bone and linearly related to the fixed depth of the screw. The load pattern of pull-out force of aDSBF model was similar to that of aSSBF model. The ultimate torque of aDSBF model was higher than that of aSSBF model, but the load pattern of ultimate torque of both models was similar to each other when the fracture reset was satisfactory, and the screw nut attaches closely to coronoid process. Moreover, with enhancement of initial pre-tightening force, the increase of ultimate torque of both models was small., In addition to three pull-out stability factors of smart screw fixations, fracture surface fitting degree and nut fitting degree are the other two important anatomical and biomechanical stability factors of smart screw fixations both for rotational stability. When all pull-out stability and rotational stability factors meet reasonable conditions simultaneously, single or double screw fixation methods are stable for the treatments of ulnar coronoid basal fractures.
资助项目National Natural Science Foundation of China [81271663, 31471146]; Wenzhou Medical University Scientific Development Foundation of Zhejiang, China [QTJ06012]; Basic Scientific Research Project of Wenzhou City of China [Y20210398]; Zhejiang Province Science and Technology Plan Research and Xinmiao Talent Program, China [2021R413036]
出版者WILEY
ISSN1757-7853
EISSN1757-7861
卷号15期号:4页码:1072-1084
DOI10.1111/os.13664
页数13
WOS类目Orthopedics
WOS研究方向Orthopedics
WOS记录号WOS:000916847200001
收录类别PUBMED ; SCIE ; SCOPUS
在线发表日期2023-01
URL查看原文
PubMed ID36647280
SCOPUSEID2-s2.0-85146656692
通讯作者地址[Wu, Lijun]Institute of Digitized Medicine and Intelligent Technology,Wenzhou Medical University,Zhejiang,325000,China
Scopus学科分类Surgery;Orthopedics and Sports Medicine
引用统计
文献类型期刊论文
条目标识符https://kms.wmu.edu.cn/handle/3ETUA0LF/172435
专题数字化医学与智能技术研究院
附属第二医院
其他_附属第三医院(瑞安市人民医院)
通讯作者Wu, Lijun
作者单位
1.Institute of Digitized Medicine and Intelligent Technology,Wenzhou Medical University,Wenzhou,China;
2.Department of Orthopedics,Tianjin Teda Hospital,Tianjin,China;
3.Department of Orthopedics,The Third Affiliated Hospital,Wenzhou Medical University,Wenzhou,China;
4.Department of Orthopedics,The Second Affiliated Hospital,Wenzhou Medical University,Wenzhou,China
第一作者单位数字化医学与智能技术研究院
通讯作者单位数字化医学与智能技术研究院
第一作者的第一单位数字化医学与智能技术研究院
推荐引用方式
GB/T 7714
Ye, Hao,Yang, Yongchao,Xing, Tingyang,et al. Anatomical and Biomechanical Stability of Single/Double Screw-Cancellous Bone Fixations of Regan-Morry Type III Ulnar Coronoid Fractures in Adults: CT Measurement and Finite Element Analysis[J]. Orthopaedic surgery,2023,15(4):1072-1084.
APA Ye, Hao., Yang, Yongchao., Xing, Tingyang., Tan, Guirong., Jin, Shuxun., ... & Wu, Lijun. (2023). Anatomical and Biomechanical Stability of Single/Double Screw-Cancellous Bone Fixations of Regan-Morry Type III Ulnar Coronoid Fractures in Adults: CT Measurement and Finite Element Analysis. Orthopaedic surgery, 15(4), 1072-1084.
MLA Ye, Hao,et al."Anatomical and Biomechanical Stability of Single/Double Screw-Cancellous Bone Fixations of Regan-Morry Type III Ulnar Coronoid Fractures in Adults: CT Measurement and Finite Element Analysis".Orthopaedic surgery 15.4(2023):1072-1084.

条目包含的文件

条目无相关文件。
个性服务
查看访问统计
谷歌学术
谷歌学术中相似的文章
[Ye, Hao]的文章
[Yang, Yongchao]的文章
[Xing, Tingyang]的文章
百度学术
百度学术中相似的文章
[Ye, Hao]的文章
[Yang, Yongchao]的文章
[Xing, Tingyang]的文章
必应学术
必应学术中相似的文章
[Ye, Hao]的文章
[Yang, Yongchao]的文章
[Xing, Tingyang]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。